ABSTRACT:
Up to 30-40% of patients undergoing cardiac resynchronization therapy (CRT) are
described as nonresponders since the initial studies. This paradigm has inspired
several modifications of the devices, electrodes and surgical technique in the
implant. The definition of CRT response should be rethought, standardized, and
ratings based on structural and/or clinical response should be proposed. The
authors discuss a series of cases in which sustained clinical improvement was
achieved despite structural worsening. Objective: To assess the profile of
clinical responders to CRT who have worsened structurally. Method: It is a
retrospective cohort of patients in outpatient follow-up from January 2012 to
March 2017. We included 13 patients (2.7%) out of a total of 476 submitted to
CRT. Inclusion criteria were to present an improvement in functional class
according to the New York Heart Association criteria (FC-NYHA) = 1
sustained for at least one year and absence of improvement or worsening of the
structural parameters evaluated by transthoracic echocardiogram [ejection
fraction (EF), diastolic diameter (LVDD) and systolic diameter (LVSD)]. The
variables analyzed were age, gender, FC-NYHA, cardiopathy, echocardiographic
and
electrocardiographic parameters, medications in use, location of implanted
electrodes, device programming, cardiary defibrillator therapies, and mortality.
Statistical analysis was performed using non-parametric Wilcoxon and McNemar
tests. Results: There were 13 patients, 92% male, mean age 60.9 ± 9.2
years and mean follow-up of 3.3 ± 1.1 years, 76% of CRT associated with
implantable cardioverter defibrillator (CRT-D). In pre-implantation, 84.6% were
in FC-NYHA III and then 61.5% were in FC-NYHA I (p = 0.001). The mean
pre-implantation EF was 31.3 ± 7.6% and 26.6 ± 7.3 (p = 0.002) in
the last evaluation. The predominant heart disease was non-ischemic in 92.5%,
most of which were chagasic cardiomyopathy (CCM) (66%). In the TRC-D group, no
shock therapy was recorded in the period; there was one death in a patient with
ischemic cardiomyopathy (IC) for the septic shock of pulmonary focus after 2.2
years of follow-up. The mean QRS was 189.9 ± 23.1 ms to 157.9 ±
35.2 after CRT (p = 0.032). There was no significant change in pre-and
post-implant medications during follow-up. Conclusion: The absence of structural
improvement should not be considered therapeutic failure, since CRT seeks to
modify the electrical activation, and may be related to better performance and
decrease of symptoms, even in evolutionary heart diseases.
Keywords: Cardiac resynchronization therapy; Cardiac insufficiency; Echocardiography; Cardiac electrophysiology.
RESUMO:
Até 30-40% dos pacientes submetidos à terapia de ressincronizaçao cardíaca (TRC)
sao descritos como nao respondedores desde os trabalhos iniciais. Esse paradigma
tem inspirado diversas modificaçoes dos dispositivos, eletrodos e técnica
cirúrgica no implante. A definiçao de resposta à TRC deverá ser repensada,
padronizada, e classificaçoes pautadas na resposta estrutural e/ou clínica devem
ser propostas. Os autores discutem uma série de casos em que se obteve melhora
clínica sustentada a despeito da piora estrutural. Objetivo: Avaliar o perfil
dos pacientes respondedores clínicos à TRC que pioraram estruturalmente. Método:
Trata-se de coorte retrospectiva de pacientes em seguimento ambulatorial de
janeiro de 2012 a março de 2017. Foram incluídos 13 pacientes (2,7%) de um total
de 476 submetidos à TRC. Os critérios de inclusao foram apresentar melhora da
classe funcional pelos critérios da New York Heart Association (CF-NYHA)
= 1 sustentada por pelo menos um ano e ausência de melhora ou com piora
dos parâmetros estruturais avaliados pelo ecocardiograma transtorácico [fraçao
de ejeçao (FE), diâmetro diastólico (DDVE) e diâmetro sistólico (DSVE)]. As
variáveis analisadas foram idade, gênero, CF-NYHA, cardiopatia, parâmetros
ecocardiográficos e eletrocardiográficos, medicaçoes em uso, localizaçao do
implante dos eletrodos, programaçao do dispositivo, terapias do
cardiodesfibrilador e mortalidade. A análise estatística foi realizada por meio
dos testes nao paramétricos de Wilcoxon e McNemar. Resultado: Foram 13
pacientes, sendo 92% do sexo masculino, idade média de 60,9 ± 9,2 anos e
seguimento médio de 3,3 ± 1,1 anos, 76% de TRC associada a
cardiodesfibrilador implantável (TRC-D). No pré-implante, 84,6% encontravam-se
em CF-NYHA III e, em seguida, 61,5% estavam em CF-NYHA I (p = 0,001). A FE média
pré-implante foi de 31,3 ± 7,6% e 26,6 ± 7,3 (p = 0,002) na última
avaliaçao. A cardiopatia predominante foi a nao isquêmica em 92,5%, sendo a
maioria cardiomiopatia chagásica (CMC) (66%). No grupo TRC-D, nao foi registrada
terapia de choque no período; houve um óbito em um paciente com cardiomiopatia
isquêmica (CMI) por choque séptico de foco pulmonar após 2,2 anos de seguimento.
O QRS médio foi de 189,9 ± 23,1 ms para 157,9 ± 35,2 após TRC (p =
0,032). Nao houve mudança significativa nas medicaçoes administradas pré- e
pós-implante durante o seguimento. Conclusao: A ausência de melhora estrutural
nao deve ser considerada falha terapêutica, pois a TRC procura modificar a
ativaçao elétrica, podendo estar relacionada a melhor desempenho e diminuiçao
dos sintomas, mesmo em cardiopatias evolutivas.
Palavras-chave: Terapia de ressincronizaçao cardíaca; Insuficiência cardíaca; Ecocardiografia; Eletrofisiologia cardíaca.
Citation: Chiarini R, Duarte CE, Silva TR, Sbaraini AB, Passuello GG, Jesus LD, et al. Cardiac Resynchronization Therapy: the Structural Response Is not Always Necessary. Arq Bras Cardiol 32(1):38. doi:10.24207/jac.v32i1.999_PT |
Received: Março 11 2019; Accepted: Maio 09 2019 |
INTRODUCTION
Cardiac pacing began to be thought of as adjuvant therapy in the treatment of cardiac insufficiency (CI) refractory to pharmacological treatment in the mid-1990s. The theoretical and experimental basis of this application is the correction of delayed left ventricular (LV) contraction in patients with IC that present complete left bundle branch block (LBBB). In these cases, with the advent of imaging exams such as echocardiography, it can be confirmed that, in addition to the delay in electrical activation, there is mechanical dyssynchrony of the ventricle, decreasing cardiac efficiency, and may promote cardiac remodeling in the short and long term. Important randomized studies supported implant guidelines adopted in Brazil and by the leading scientific societies in the United States and Europe (American College of Cardiology (ACC), American Heart Association (AHA), Heart Rhythm Society (HRS) and European Society of Cardiology (ESC)], being recommended as a recommendation class I and level A evidence of cardiac resynchronization therapy (CRT) for NYHA II and III patients, QRS duration = 150 ms and LBBB, provided that with optimized drug therapy, sinus rhythm and EF = 35%8-11. Although there are no important randomized studies, it is now possible to identify other groups of patients with a high probability of positive response to CRT, setting guidelines as indication class I or II with a level of evidence B8-11. Despite the technical advances of devices, electrodes, implant surgery, and development of complementary methods, more recent studies still classify about 30% of patients as non-responders to CRT despite the lack of consensus in the definition of CRT response1-8. The present study analyzed a retrospective cohort of patients submitted to CRT, evaluating the profile of clinical responders to therapy, seeking to identify factors associated with the mechanism of action that results in clinical improvement without the presence of structural improvement.
METHODS
Retrospective study of a cohort of follow-up patients at the Advanced Center of Ritology and Electrophysiology (Centro Avançado de Ritmologia e Eletrofisiologia-CARE), in the state of Sao Paulo, submitted to the CRT from January 2012 to March 2017.
Age, gender, FC-NYHA, cardiopathy, echocardiographic parameters, and pre and post-CRT electrocardiographic parameters, medications in use, electrode implant localization, device programming, cardi-defibrillator therapies, and mortality were analyzed. The criteria for an indication of CRT and outpatient follow-up obeyed the institutional protocol that follows the Brazilian Guidelines for Implantable Electronic Cardiac Devices.
The criterion used to define exclusively a clinical response to CRT was: improvement of the SF by NYHA criteria = 1 sustained for a minimum period of 12 months after CRT, without improvement or worsening of the structural parameters in the transthoracic echocardiogram.
Exclusion criteria were any increase in left ventricular ejection fraction (EF) and/or reduction of diastolic and/or final systolic diameters evaluated by transthoracic echocardiography or unsupported clinical improvement for a minimum period of 12 months.
Statistical analysis
Continuous variables were expressed in means ± standard deviation (SD) and categorical variables in percentages. Statistical analysis was performed using non-parametric Wilcoxon and McNemar tests. The representation was made by the mean, SD and quartiles (Tables 1, 2 and 3).
Medication | Pre-TRC | Post-TRC | p-value |
---|---|---|---|
n (%) | |||
Angiotensin converting enzyme inhibitors | 6 (46.2) | 7 (53.8) | 0.317 |
Angiotensin receptor blockers | 5 (38.5) | 4 (30.8) | 0.317 |
Furosemide | 8 (61.5) | 7 (53.8) | 0.564 |
Beta blocker | 13 (100.0) | 13 (100.0) | - |
Spironolactone | 10 (76.9) | 9 (69.2) | 0.564 |
Amiodarone | 9 (69.2) | 10 (76.9) | 0.564 |
Propafenone | 1 (7.7) | 0 (0.0) | 1.000 |
Nitrate | 1 (7.7) | 1 (7.7) | - |
Digoxin | 2 (15.4) | 2 (15.4) | - |
Acetylsalicylic acid | 4 (30.8) | 4 (30.8) | - |
Oral anticoagulant | 5 (38.5) | 3 (23.1) | 0.157 |
Statin | 8 (61.5) | 8 (61.5) | - |
Variiable | Moment | n | Mean | DP | Minimum | Maximum | P25 | Medium | P75 | p-value |
---|---|---|---|---|---|---|---|---|---|---|
Pre-TRCPost-TRC | 1010 | 32.3031.60 | 2.872.80 | 27.0026.00 | 37.0036.00 | 30.7529.75 | 32.0032.00 | 34.5033.25 | 0.471 | |
Pre-TRCPost-TRC | 1212 | 45.7548.00 | 10.209.83 | 32.0035.00 | 64.0070.00 | 35.7539.00 | 46.0046.00 | 55.0054.50 | 0.167 | |
Pre-TRCPost-TRC | 77 | 53.2963.29 | 5.196.82 | 46.0050.00 | 59.0072.00 | 49.0061.00 | 55.0064.00 | 58.0067.00 | 0.018 | |
Pre-TRCPost-TRC | 1313 | 65.5470.38 | 5.977.50 | 58.0054.00 | 81.0084.00 | 61.0067.00 | 65.0070.00 | 68.5076.50 | 0.015 | |
Pre-TRCPost-TRC | 1111 | 8.188.36 | 1.081.75 | 7.006.00 | 10.0011.00 | 7.007.00 | 8.008.00 | 9.0010.00 | 0.732 | |
Pre-TRCPost-TRC | 1111 | 8.188.00 | 1.171.34 | 7.006.00 | 10.0011.00 | 7.007.00 | 8.008.00 | 9.009.00 | 0.726 | |
Pre-TRCPost-TRC | 1313 | 31.3126.62 | 7.937.63 | 18.0015.00 | 46.0041.00 | 28.0019.50 | 32.0028.00 | 37.0031.00 | 0.002 | |
Pre-TRCPost-TRC | 33 | 32.3329.00 | 0.588.19 | 32.0022.00 | 33.0038.00 | 32.0022.00 | 32.0027.00 | 33.0038.00 | - | |
QRS (ms) | Pre-TRCPost-TRC | 1313 | 189.92157.92 | 29.1035.21 | 150.0080.00 | 240.00200.00 | 160.00127.50 | 190.00160.00 | 204.50180.00 | 0.032 |

SâQRS | Pre-TRC | Post-TRC | p-value |
---|---|---|---|
n (%) | |||
I (-90 to 0º) | 7 (53.8) | 1 (7.7) | 0.027 |
II (0 to +90º) | 2 (15.4) | 4 (30.8) | |
III (+90 to 180º) | 3 (23.1) | 6 (46.2) | |
IV (180 to -90º) | 1 (7.7) | 2 (15.4) |
RESULTS
A total of 476 sequential charts of patients submitted to CRT-related surgical intervention from January 2012 to March 2017 were selected. Thirteen patients (2.7%) met the inclusion criteria, most were male (92%), with a mean age of 60.9 ± 9.2 years. Before implantation, FC-NYHA was III in 84.6%, IV in 7.6% and II in 7.6%; Mean EF of the LV was 31.3 ± 7.9%, the mean systolic diameter (LVSD) was 53.2 ± 5.1 mm and the mean diastolic diameter (LVDD) was 65.5 ± 5.9. Baseline heart disease was distributed in: Chagas cardiomyopathy (CCC) in 61%; hypertensive (HiCM) in 15%; dilated (DCM) in 8%; ischemic (ICM) in 8%; and hypertrophic (HCM) in 8%. Pre-implantation sinus rhythm was present in 76% of the patients, with a mean QRS of 189.9 ± 29.1 ms, with an LBBB pattern being identified in 61.5%, of which 25% were associated with axis deviation for left. Right bundle branch block (RBBB) was evident in 38.8%, 60% of which were associated with an anterosuperior divisional block (ADB) and 20% associated with a posteroinferior divisional block (PIDB). The median pre-implant axis was between -90 to 0º in 53%. Most were using pharmacological therapy at optimized doses pre- and post-implant of the device. There were no significant changes in the disease-modifying drugs, such as beta-blockers, angiotensin converting enzyme inhibitor (ACEI), or potassium-sparing angiotensin II receptor blocker (ARB) and diuretic with pre-and post-CRT 100 and 100%, 84.6 and 84.6%, 76.9 and 69.2%, respectively. The loop diuretic rate remained at 54% in the pre- and post-implant periods (Table 1). After CRT, with a mean follow-up of 3.3 ± 1.1 years, there was the improvement of at least 1 point in NYHA-FC in 100% of the patients (p = 0.001) and in 62% there was the improvement of 2 points. The mean EF was 31.3 ± 7.9% in the pre-implant period to 26.6 ± 7.6% post-implant (p = 0.002). There were increased diameters after CRT: mean LVSD 63.2 ± 6.8 mm (p = 0.018) and mean LVDD 70.3 ± 7.5 mm (p = 0.015).
The devices were programmed in 76% in the DDD mode, being 53% multipoint, with an average atrioventricular (AV) interval of 214 ms. The biventricular pacing rate was 98.7%, with the LV-RV pacing pattern 61.5%, RV-LV 23%, and simultaneous 15.5%. The QRS under biventricular pacing was 187.1 ± 25.9 ms with ADB in 7% and PIDB in 46%, mean axis under stimulation between +90 and + 180º in 43% and between 0 and + 90º in 30%. Regarding the position of the electrodes, the right ventricle was superior to the septal-basal, super-septal-medial and mid-septal-apical positions in 38, 23 and 38% of the implants, respectively. The LV electrode in mid-latero-basal position in 23%, supero-latero-basal 15%; medial-latero-medial 15% infero-latero-apical 15%, infero-latero-medial 15% and superior-antero-basal 15%. There were no records of cardiac defibrillator shock therapies. There was one death in a patient with IMC related to pulmonary focus septic shock with 2.2 years of CRT (Tables 1 and 2).
DISCUSSION
This study discusses the standardization of CRT response criteria and proposes analysis from both the structural and clinical perspectives, questioning the real need to restrict the indication of CRT to those who have established predictors of structural response, such as the presence of LBBB and QRS> 150 ms in patients with DCM.
It should be remembered that in the onset of CRT, with the study MUSTIC15 was analyzed exactly that the functional capacity was improved, the quality of life and the patient>s preference for the mode of stimulation, with LV stimulation bound or not. Restricting CRT to the point of structural improvement would deny its greatest benefit to those with evolutionary structural cardiomyopathy.
Question: Can driving disorders be the cause of CI or just an aggravating factor? In the MADIT-CRT study25, in women with mild IC symptoms, female gender, absence of infarction, LBBB, QRS> 150 ms, body mass index (BMI) <30 kg / m2 and reduced left atrial volume were predictors of normalization of parameters and clinical improvement before CRT (super-responders). In these cases CRT is very likely to act on the pathophysiology of ventricular dysfunction, restoring normalization.
In this context, we discussed whether the clear change in the direction of activation (Table 3) and the search for correction of conduction disturbance were determinants of improvement in clinical performance even in those whose etiology was of evolutionary cardiomyopathies, such as Chagas> disease.
The major criticism of these results is the deficiency of other parametric clinical evaluations. On MUSTIC15, in addition to FC-NYHA, the 6-minute walk test, the peak O2 consumption (VO2 peak), and the Minnesota quality of life score were analyzed. Clinical evaluation of this study was limited to FC-NYHA. Regarding echocardiographic references, a deficiency is present in the evaluation of pre-and post-implant mitral regurgitation, since it could be a determinant of response. It is believed that the reduction of mitral regurgitation before CRT may lead to increased pressure load in the LV, resulting in increased diameters with worsening of global EF, but generating a greater anterograde flow through the LV exit pathway, resulting in clinical improvement. The actual understanding of the clinical improvement of this group of patients should encourage prospective studies to study this effect.
CONCLUSION
The absence of structural improvement should not be considered therapeutic failure. Electrical dyssynchrony can lead to changes in the sense of activation, causing slower and disharmonious contractions, resulting in hemodynamic performance deficit. The CRT seeks to accelerate and harmonize electrical activation, correcting driving disorders that worsen cardiac performance, and may promote clinical and/or structural improvement.
REFERENCES






















